Use of "Quick-Disconnect" Fittings Results in Laboratory Instrument Explosion

Use of "Quick-Disconnect" Fittings Results in Laboratory Instrument Explosion

Incident Date
When Incident Was Discovered
Severity
Incident
Was Hydrogen Released?
Yes
Was There Ignition?
Yes
Ignition Source
spark from switching instrument on
Setting
Damage and Injuries
Contributing Factors
Describe the incident, including corrective steps taken and their result.

A researcher was using numerous compressed gases in his lab. To facilitate reconfiguring his experimental apparatus, he installed "quick-disconnect" fittings on flexible tubing connected to his compressed gas cylinders/regulators. He also fitted all of the equipment that needed gas with complementary "quick-disconnect" fittings.

The day of the incident, he needed to purge his IR spectrometer with nitrogen as the element heated up. He mistakenly attached the "quick-disconnect" fitting from a cylinder of 10% nitrogen and 90% hydrogen to the "quick-disconnect" fitting on his spectrometer. As soon as the gas started flowing and he switched on the element, the instrument exploded, completely destroying a $6,000 piece of equipment. Only minor damage was done to the rest of the room, and luckily there were no injuries.

Lessons Learned

The lessons learned from this incident are:

  1. Verify the gas that you are using.
  2. Avoid using "quick-disconnect" fittings in this type of situation. If they are absolutely needed, there are sets available that ensure that every pair of matching fittings are different from all the others.

Key:

  • = No Ignition
  • = Explosion
  • = Fire
Hydrogen Incident Summaries by Equipment and Primary Cause/Issue
Equipment / CauseEquipment Design or SelectionComponent FailureOperational ErrorInstallation or MaintenanceInadequate Gas or Flame DetectionEmergency Shutdown ResponseOther or Unknown
Hydrogen Gas Metal Cylinder or Regulator 3/31/2012
4/30/1995
2/6/2013
4/26/201012/31/1969  3/17/1999
11/1/2001
12/23/2003
Piping/Valves4/4/2002
2/2/2008
5/11/1999
4/20/1987
11/4/1997
12/31/1969
8/19/1986
7/27/1991
12/19/2004
2/6/2008
10/3/2008
4/5/2006
5/1/2007
9/19/2007
10/31/1980
2/7/20091/24/1999
2/24/2006
6/8/1998
12/31/1969
2/7/2009
9/1/1992
10/31/1980
10/3/2008 
Tubing/Fittings/Hose 9/23/1999
8/2/2004
8/6/2008
9/19/2007
1/1/19829/30/2004
10/7/2005
 10/7/2005 
Compressor 10/5/2009
6/10/2007
8/21/2008
1/15/2019
  10/5/20098/21/2008 
Liquid Hydrogen Tank or Delivery Truck4/27/198912/19/2004
1/19/2009
8/6/200412/31/1969 1/1/197412/17/2004
Pressure Relief Device7/25/2013
5/4/2012
1/15/2002
1/08/2007
12/31/1969    
Instrument1/15/20193/17/1999
12/31/1969
2/6/2013
  11/13/73  
Hydrogen Generation Equipment7/27/1999  10/23/2001   
Vehicle or Lift Truck 7/21/2011    2/8/2011
12/9/2010
Fuel Dispenser 8/2/2004
5/1/2007
6/11/2007
9/19/2007
 2/24/2006
1/22/2009
   
Fuel Cell Stack      5/3/2004
12/9/2010
2/8/2011
Hydrogen Cooled Generator   12/31/1969
2/7/2009
   
Other (floor drain, lab
anaerobic chamber,
heated glassware,
test chamber,
gaseous hydrogen
composite cylinder,
delivery truck)
 11/14/1994
7/21/2011
7/27/1999
6/28/2010
8/21/2008
12/31/1969
3/22/2018
  6/10/2019
  • = No Ignition
  • = Explosion
  • = Fire