Hydrogen Explosion in Microbiological Anaerobic Chamber

Hydrogen Explosion in Microbiological Anaerobic Chamber

When Incident Was Discovered
Severity
Incident
Was Hydrogen Released?
Yes
Was There Ignition?
Yes
Setting
Contributing Factors
Describe the incident, including corrective steps taken and their result.

An explosion occurred in a Microbiological Anaerobic Chamber of approximately 2 m3 capacity that contained an explosive mixture of hydrogen and air. A fire followed the explosion, but was rapidly extinguished by staff using fire extinguishers prior to the arrival of fire service personnel. The pressure wave from the explosion blew windows out of the laboratory, with glass hitting a passerby on a path outside and glass shards landing up to 30 m away. Ceiling panels were dislodged in the laboratory and adjacent rooms, and a worker using the apparatus at the time was taken to the hospital by ambulance for burn treatment. The worker subsequently fully recovered. Another worker in the lab at the time required medical observation, but was otherwise unharmed.

Mixtures of inert gases and hydrogen are routinely used in the type of anaerobic chamber involved in the incident. The mixtures used in the chamber involved were produced locally in the laboratory using nitrogen, carbon dioxide, and hydrogen. The hydrogen in the mixture reacts with any oxygen present in the chamber, on a heated catalyst, to eliminate oxygen and keep the chamber anaerobic. The local operating procedures used in the lab allowed high concentrations of hydrogen to be introduced into the chamber. A worker inadvertently admitted air to the chamber while doing maintenance, allowing the hydrogen-enriched atmosphere in the chamber to mix with air and subsequently ignite, most likely on contact with the oxidation catalyst in the chamber, resulting in the explosion and fire.

Hydrogen gas has a very wide range of flammability when mixed with air (4–75%). Oxidation catalysts can ignite explosive gas mixes without heating, spark, or flame. Local operating procedures and practices varied from the manufacturer’s advice. An unknown concentration of hydrogen was present in the chamber, resulting in a significant fire and explosion risk.

Lessons Learned

The practice of making gas mixtures in the laboratory should be eliminated, and gas mixtures with a known low hydrogen concentration should be purchased for use. The concentration of hydrogen used should be such that it is not possible to form an explosive mixture on dilution with air (i.e., hydrogen concentration less than 4% after mixing with air from a leak, damage to the chamber, or inadvertent admission of air to the chamber). A gas monitor with a built-in alarm should be purchased and installed to continuously monitor both hydrogen and oxygen concentrations in the chamber, and to provide visible and audible indication of any problems. The manufacturer’s instructions and manuals should be closely followed. All users should be fully trained in the use of the equipment, and should be fully conversant with the potential hazards and how to manage the associated risks. Hydrogen use in anaerobic chambers is discussed in the Lessons Learned Corner on this website.

Key:

  • = No Ignition
  • = Explosion
  • = Fire
Hydrogen Incident Summaries by Equipment and Primary Cause/Issue
Equipment / CauseEquipment Design or SelectionComponent FailureOperational ErrorInstallation or MaintenanceInadequate Gas or Flame DetectionEmergency Shutdown ResponseOther or Unknown
Hydrogen Gas Metal Cylinder or Regulator 3/31/2012
4/30/1995
2/6/2013
4/26/201012/31/1969  3/17/1999
11/1/2001
12/23/2003
Piping/Valves4/4/2002
2/2/2008
5/11/1999
4/20/1987
11/4/1997
12/31/1969
8/19/1986
7/27/1991
12/19/2004
2/6/2008
10/3/2008
4/5/2006
5/1/2007
9/19/2007
10/31/1980
2/7/20091/24/1999
2/24/2006
6/8/1998
12/31/1969
2/7/2009
9/1/1992
10/31/1980
10/3/2008 
Tubing/Fittings/Hose 9/23/1999
8/2/2004
8/6/2008
9/19/2007
1/1/19829/30/2004
10/7/2005
 10/7/2005 
Compressor 10/5/2009
6/10/2007
8/21/2008
1/15/2019
  10/5/20098/21/2008 
Liquid Hydrogen Tank or Delivery Truck4/27/198912/19/2004
1/19/2009
8/6/200412/31/1969 1/1/197412/17/2004
Pressure Relief Device7/25/2013
5/4/2012
1/15/2002
1/08/2007
12/31/1969    
Instrument1/15/20193/17/1999
12/31/1969
2/6/2013
  11/13/73  
Hydrogen Generation Equipment7/27/1999  10/23/2001   
Vehicle or Lift Truck 7/21/2011    2/8/2011
12/9/2010
Fuel Dispenser 8/2/2004
5/1/2007
6/11/2007
9/19/2007
 2/24/2006
1/22/2009
   
Fuel Cell Stack      5/3/2004
12/9/2010
2/8/2011
Hydrogen Cooled Generator   12/31/1969
2/7/2009
   
Other (floor drain, lab
anaerobic chamber,
heated glassware,
test chamber,
gaseous hydrogen
composite cylinder,
delivery truck)
 11/14/1994
7/21/2011
7/27/1999
6/28/2010
8/21/2008
12/31/1969
3/22/2018
  6/10/2019
  • = No Ignition
  • = Explosion
  • = Fire