Hydrogen Gas Leak from Compressed Gas Cylinder
Two scientists were changing hydrogen gas cylinders in an analytical laboratory. They were in the process of removing the cylinder cap from the new cylinder when a loud hissing noise occurred and they quickly realized that the tank was leaking. After making a quick attempt to shut off the tank, which was not possible, they left the lab and notified their supervisor.
After checking that everyone was out of the lab, the supervisor paged all staff in the vicinity to immediately evacuate to the staging area. Facility management and ES&H management were notified about the situation, and they contacted the local fire department to respond to the site in case the venting gas was ignited.
The first responders arrived quickly and spoke with facility management and the site safety officer. Everyone agreed that the best course of action was to allow the tank to vent until it was empty. The hydrogen gas supplier was called, but a specialist was unavailable at that time. Based on information from the hydrogen supplier's website and some basic estimates of the hydrogen release rate, facility staff determined that the contents of the tank should have been completely vented. A decision was made that the first responders would approach the lab in full bunker gear with SCBA and a gas detector (Lower Explosion Limit (LEL) meter) to evaluate if there was any gas present. If they heard any hissing coming from the tank or if there was any reading above "0%" on the LEL meter, the team would immediately pull back and reassess the situation.
Two first responders entered the lab and two were stationed as back-up. Upon entry to the lab, the first responders indicated that they did not hear any hissing and proceeded towards the cylinder with the LEL meter. Their readings confirmed that the cylinder had vented and the lab was clear and safe to enter. The hydrogen cylinder was moved outside and secured and later returned to the supplier.