Hydrogen Balloon Incident

Hydrogen Balloon Incident

Incident Date
When Incident Was Discovered
Severity
Incident
Was Hydrogen Released?
Yes
Was There Ignition?
Yes
Damage and Injuries
Describe the incident, including corrective steps taken and their result.

During the early morning hours on a Tuesday, a university support staff member was preparing for an off-campus community outreach program for high-school-age students in the community. One of the program demonstrations was to show students the reaction energy and properties of the hydrogen + oxygen = water chemical reaction. It was a demonstration that the professor and the staff member overseeing the program had done for over 15 years with no incidents ever occurring.

To prepare for the demonstration, eight balloons were filled, four with pure hydrogen and four with the proper combination of hydrogen and oxygen. The balloons were placed into a larger plastic garbage bag and carried outside to a university-owned SUV located next to the building's loading dock. The bag of balloons was being placed into the vehicle from the right side door. The balloons were inside the vehicle, but not yet in their final place for transport. As the staff member was leaning into the doorway during this placement, a balloon at the top end of the bag ignited, which then set off a fiery explosive reaction involving the other balloons.

The staff member, who received burns to the face, was transported to a nearby hospital by the professor. The vehicle's windshield and two back windows were damaged, and the headliner of the vehicle was slightly scorched.

Lessons Learned

Although the preparation-for-transport procedures were done the same way they were done for previous outreach programs, this time it proved to be a different situation. It is not clear what caused the ignition of the first balloon, which then set off a chain reaction to the others. The incident shows that preparation for transport is a very important element in the overall process, and it should be evaluated for risk factors along with every other element of the process.

Key:

  • = No Ignition
  • = Explosion
  • = Fire
Hydrogen Incident Summaries by Equipment and Primary Cause/Issue
Equipment / CauseEquipment Design or SelectionComponent FailureOperational ErrorInstallation or MaintenanceInadequate Gas or Flame DetectionEmergency Shutdown ResponseOther or Unknown
Hydrogen Gas Metal Cylinder or Regulator 3/31/2012
4/30/1995
2/6/2013
4/26/201012/31/1969  3/17/1999
11/1/2001
12/23/2003
Piping/Valves4/4/2002
2/2/2008
5/11/1999
4/20/1987
11/4/1997
12/31/1969
8/19/1986
7/27/1991
12/19/2004
2/6/2008
10/3/2008
4/5/2006
5/1/2007
9/19/2007
10/31/1980
2/7/20091/24/1999
2/24/2006
6/8/1998
12/31/1969
2/7/2009
9/1/1992
10/31/1980
10/3/2008 
Tubing/Fittings/Hose 9/23/1999
8/2/2004
8/6/2008
9/19/2007
1/1/19829/30/2004
10/7/2005
 10/7/2005 
Compressor 10/5/2009
6/10/2007
8/21/2008
1/15/2019
  10/5/20098/21/2008 
Liquid Hydrogen Tank or Delivery Truck4/27/198912/19/2004
1/19/2009
8/6/200412/31/1969 1/1/197412/17/2004
Pressure Relief Device7/25/2013
5/4/2012
1/15/2002
1/08/2007
12/31/1969    
Instrument1/15/20193/17/1999
12/31/1969
2/6/2013
  11/13/73  
Hydrogen Generation Equipment7/27/1999  10/23/2001   
Vehicle or Lift Truck 7/21/2011    2/8/2011
12/9/2010
Fuel Dispenser 8/2/2004
5/1/2007
6/11/2007
9/19/2007
 2/24/2006
1/22/2009
   
Fuel Cell Stack      5/3/2004
12/9/2010
2/8/2011
Hydrogen Cooled Generator   12/31/1969
2/7/2009
   
Other (floor drain, lab
anaerobic chamber,
heated glassware,
test chamber,
gaseous hydrogen
composite cylinder,
delivery truck)
 11/14/1994
7/21/2011
7/27/1999
6/28/2010
8/21/2008
12/31/1969
3/22/2018
  6/10/2019
  • = No Ignition
  • = Explosion
  • = Fire