Gas Chromatograph Piping Leak Explodes in Instrument Room
An explosion occurred at a chemical plant in an analysis room containing various analyzer instruments, including a gas chromatograph supplied with hydrogen. A contract operator was performing work to install a new vent line to a benzene analyzer that was part of a group of CO2 analyzers, but separate and unrelated to the gas chromatograph. During the process of this work, a plant supervisor accompanying the contract operator doing the work had an indication of flammable gas present on a portable detector. This was in conflict with the fixed gas detector in the analysis room that was indicating that no flammable gas was present. As a precaution, the plant supervisor immediately cut off the hydrogen supply and, along with the contract operator, began the normal task of determining if there was a gas leak in the room. While investigating the potential gas leak, the gas chromatograph that was housed in a cabinet exploded and the nearby contract operator was injured in the chest and hand when the cabinet door burst open and received minor burns to the hair and face from the fire. See Attachments 1 and 2 for diagrams and photos.
The plant control room was immediately notified by radio of the incident and subsequently called emergency services to send an ambulance to treat the injured contract operator. The ambulance arrived at the plant in approximately 15 minutes and, as a precautionary measure, the contract operator was transported to a nearby hospital for further assessment. There was no internal damage or damage to the airways and no additional medical injuries were found. Minor damage to the gas chromatograph and cabinet occurred from the explosion/fire.
The flame ionization gas chromatograph (used in this application to measure the quality of specific hydrocarbon) was installed in a small cabinet that was not rated for explosive atmosphere application, and the natural ventilation through cabinet louvers was insufficient. In addition, the cabinet louvers were not connected to an exhaust duct, allowing gases to escape into a room not rated for explosive atmosphere. The analysis room's fixed gas detector was located too far from the cabinet and was not mounted high enough to detect the small gas leak from the gas chromatograph system. The valve that shuts off the hydrogen supply was located near the gas chromatograph's inlet, and it would be better located outside of the analysis room. The origin of the leak has not yet been pinpointed, but the explosion occurred either in the gas chromatograph or in its cabinet. These deficiencies have since been corrected and are shown in the diagrams presented in Attachment 1. Photos of the gas chromatograph and associated cabinet are shown in Attachment 2.