Explosion at a bio-fuels research laboratory
The incident at a bio-fuels research laboratory occurred when a visiting research fellow was transferring hydrogen, oxygen and carbon dioxide from compressed gas cylinders into a low-pressure mixing tank. The mixture, consisting of 70% H2, 25% O2 and 5% CO2, was used as a bacterial growth medium. The research fellow had just finished to fill the tank when an explosion occurred. The explosion was the result of leaks which allowed the formation of a flammble and explosive mixture, sparked probably by the use of an inproper pressure gauge or by static electricity. The researcher was severly injured by the explosion.
Event Initiating System
Classification of the Physical Effects
Nature of the Consequences
Cause Comments
The INITATING CAUSE was the creation of a flammable mixture in the surrounding of the tank due to leak, ignited by a spark of the pressure gauge, possibly assisted by static electricity shock. All filling operations, including their controls, were performed manually. Except for a measure of the internal tank total pressure by means of a portable pressure gauge, no other monitoring system was used, despite the specific hazards related to handling hydrogen and oxygen. In particularly, no flammable gas detector is mentioned having been used. Moreover, leak tests after assembly had not been performed at the nominal (rated) tank pressure, but at less than 2/3 of it. The ROOT CAUSE was related to lack of proper risk assessment, lack of preventing and mitigating safety systems, and lack of duly supervision by university management of the activities if a visiting fellow.