Explosion in an analytical laboratory (766)

This data is from the European Hydrogen Incidents and Accidents database HIAD 2.1, European Commission, Joint Research Centre.

Explosion in an analytical laboratory

An explosion and fire occurred in a mud logging unit during the final phase of rigging it up. Gas analysers had been calibrated and the flame ionization detection (FID) and total hydrocarbon analyser (THA) modules were left in operation so that they could stabilize. The mud logging personnel were connecting the outside sample points and drains when the explosion and fire occurred. The mud logging unit used FID instrumentation that required hydrogen gas to be piped into the unit to serve as the fuel for the flame within the instrumentation. The piping and hose connections used to supply the hydrogen gas to the unit were neither pressure tested nor checked for leaks; the unit had no gas detection equipment installed to detect either hydrogen or test/sample gas, nor did the hydrogen bottle have any shutdown equipment installed to prevent flow should a leak occur. In addition, the inside connection to the detection modules consisted of polyurethane hose, which was either blown loose or burned off the connector during the initial explosion and fire. This loss of confinement allowed the hydrogen to continue to flow and feed the fire in the mud logging unit.
Event Date
July 22, 1998
Record Quality Indicator
Event Initiating System
Classification of the Physical Effects
Nature of the Consequences
Cause Comments
The direct cause of the leak is unclear, although the lack of regular inspection and of emergency control system paid a role The root cause is absence of a risk assessment and of a safety design
Facility Information
Application Type
Application
geology
Specific Application Supply Chain Stage
Components Involved
Flame ionisation detector, hydrogen supply line
Location Type
Location description
DESCRIPTION OF THE TERMS: Mud logging = creation of a detailed record of a borehole by examining the cuttings of rock brought to the surface by the circulating drilling medium. FID = A flame ionization detector is a scientific instrument that measures analyses in a gas stream.
Operational Condition
Consequences
Number of Injured Persons
0
Number of Fatalities
0
Post-event Summary

Unknown consequence, although the fact that manual operation by operators were performed when the fire and explosion took place suggest injuries and the total destruction of the analytical equipment.

Event Nature
Release Type
Release Substance
H2 100%
Hydrogen Release Concentration (%)
0.00
Detonation
No
Deflagration
No
High Pressure Explosion
No
High Voltage Explosion
No
References
References

OCS BBS Website on Outer Continental Shelf tracking critical information on the Gulf of Mexico: Available at https://ocsbbs. com/SafetyAlerts/Item/176(accessed June 2020)

Alert report by the U. S. Department of the InteriorMinerals Management ServiceGulf of Mexico OCS Region. available at https://www. bsee. gov/sites/bsee. gov/files/safety-alerts/safety/safety-alert-no-176. pdf (accessed June 2020)