Explosion in an analytical laboratory
An explosion and fire occurred in a mud logging unit during the final phase of rigging it up. Gas analysers had been calibrated and the flame ionization detection (FID) and total hydrocarbon analyser (THA) modules were left in operation so that they could stabilize. The mud logging personnel were connecting the outside sample points and drains when the explosion and fire occurred. The mud logging unit used FID instrumentation that required hydrogen gas to be piped into the unit to serve as the fuel for the flame within the instrumentation. The piping and hose connections used to supply the hydrogen gas to the unit were neither pressure tested nor checked for leaks; the unit had no gas detection equipment installed to detect either hydrogen or test/sample gas, nor did the hydrogen bottle have any shutdown equipment installed to prevent flow should a leak occur. In addition, the inside connection to the detection modules consisted of polyurethane hose, which was either blown loose or burned off the connector during the initial explosion and fire. This loss of confinement allowed the hydrogen to continue to flow and feed the fire in the mud logging unit.
Event Initiating System
Classification of the Physical Effects
Nature of the Consequences
Cause Comments
The direct cause of the leak is unclear, although the lack of regular inspection and of emergency control system paid a role The root cause is absence of a risk assessment and of a safety design