Hydrogen leak in a chemical laboratory (697)

This data is from the European Hydrogen Incidents and Accidents database HIAD 2.1, European Commission, Joint Research Centre.

Hydrogen leak in a chemical laboratory

Hydrogen as carrier gas was provided to a reactor gas chromatograph by several compressed hydrogen bottles. The bottles emptied rapidly, but the loss of carrier gas could only be detected when the gas storage pressure fell to a level to stop the flow to the chromatographer. Later on, the leak was identified on the gas connection.

Event Date
January 7, 1990
Record Quality Indicator
Event Initiating System
Classification of the Physical Effects
Nature of the Consequences
Cause Comments

The INITIATING CAUSE was an instrumentation connection with a 'serious leak'. A ROOT CAUSE is not given. However, the failing to detect hydrogen and the quick emptying of the bottles highlights a lack of safety design of the laboratory and of effective safety procedures.

Facility Information
Application Type
Application
chemical laboratory
Specific Application Supply Chain Stage
Components Involved

chromatographer, compressed hydrogen storage

Storage/Process Medium
Location Type
Operational Condition
Consequences
Number of Injured Persons
0
Number of Fatalities
0
Lessons Learned

In this event, a non-negligible amount of hydrogen was released in the laboratory without further consequences. A chromatographer using hydrogen gas as a carrier is usually equipped with an internal safety sensor. However, this leak occurred outside on the line to the chromatograph housing, so that the internal gas detector could not detect the leak. Nothing is known on the laboratory assets and its safety measures, such as the presence of forced ventilation, fume hoods and hydrogen safety sensors. If the latter were deployed, they were clearly not in the right place to detect this leak. It can also be assumed that hydrogen could have formed a flammable mixture inside the laboratory and that an ignition source could have more serious consequences. The fact that the leak was not detected until the storage was almost empty implies a lack of effective hydrogen detection in the laboratory. Moreover, a quick emptying of compressed hydrogen gas bottles should have been easily notices by the laboratory personnel, if the right procedure would have been in place and followed.

Event Nature
Release Type
Release Substance
H2 100%
Hydrogen Release Concentration (%)
0.00
Ignition Source
References
References

Event of the UK database ICHEME. The ICHEME database is no longer available for purchase, but data can be download as PDF for free. https://www. icheme. org/knowledge/safety-centre/resources/accident-data/ (accessed October 2020)