Explosion in Microbiological Anaerobic Chamber -
The Microbiological Anaerobic Chamber of approximately 2 m3 capacity was containing an explosive mixture of hydrogen and air. A fire followed the explosion, but was rapidly extinguished by staff using fire extinguishers, prior to the arrival of fire service personnel. The pressure wave from the explosion damaged neighbouring building and an injury. The local operating procedures used in the lab allowed high concentrations of hydrogen to be introduced into the chamber. A worker inadvertently admitted air to the chamber whilst undertaking maintenance, allowing the hydrogen enriched atmosphere in the chamber to mix with air, and subsequently ignite, most probably on contact with the oxidation catalyst in the chamber, resulting in the explosion and subsequent fire.
Event Initiating System
Classification of the Physical Effects
Nature of the Consequences
Causes
Cause Comments
The INITIATING cause is related to a human error: a worker inadvertently admitted air to the chamber while doing maintenance, allowing the hydrogen-enriched atmosphere in the chamber to mix with air and subsequently ignite, most likely on contact with the oxidation catalyst in the chamber, resulting in the explosion and fire.